DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 30 Section 8-4 - Loading / Reloading / Upgrading the Software
8-4-9 Loading only the application software (continued)
Loading only application software
Table 8-12 Loading Only Application Software
Step Corresponding Graphic
1.
Turn on the LOGIQ Totus and wait until Windows has booted
and is loading. Watch carefully for the screen swirling circles,
as shown.
2.
Promptly insert the software thumb drive into the USB port on
the back of the LOGIQ Totus.
This action will trigger the application only installation. If you
wait too long, the application will start without detecting the
thumb drive and you will have start the process again.
3.
At the Start Application dialog, select Install SW....
4.
At the StartLoader dialog, select OK.