GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-4 - Loading / Reloading / Upgrading the Software 8 - 31
5.
Select OK to continue with the application software load
process.
6.
If prompted, select Install to begin the software installation.
7.
If you get an error stating “This package cannot be installed.
The package is not compatible or has been tampered. Please
contact GE HealthCare Service”, select OK and try to select the
correct file again.
8.
Once the application software installation is complete this
screen will display, remove the thumb drive before the system
re-starts.
The system will boot up.
9.
Let the load process continue. It will take several minutes. DO
NOT disturb the process.
Table 8-12 Loading Only Application Software (Continued)
Step Corresponding Graphic