GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 32 Section 8-4 - Loading / Reloading / Upgrading the Software
10.
Once the application software installation is complete this
screen will display, remove the thumb drive before the system
re-starts.
The system will boot up.
11.
After boot up, log in as ADM (no password required) and, when prompted, select Lowest for the security level.
NOTE: Select the Lowest security level and then continue with installation and testing. After the update is complete, you can
transfer the LOGIQ Totus to the Site Administrator who can reset the security level for the site. The security level can be reset
through Utility > Admin > Logon.
12.
Whether encryption is On or Off and after a load of the
application software (only), the New Software Verification
dialog box may display outlining the functional checks required
for the software installation.
Select each of the functional checks.
Enter a signature and select OK.
Table 8-12 Loading Only Application Software (Continued)
Step Corresponding Graphic