8 - 36 Section 8-4 - Loading / Reloading / Upgrading the Software
8-4-10 Loading software from the internal partition (continued)
Verifying the software versions:
1.) From the touch panel, select Utility > System > About.
2.) Verify software version matches the version in the software used during installation, version
received with kit.
Check patient archive
Confirm that the patient archive is available. If the patient archive is encrypted, you will need the
encryption key or password to unlock.
Functional checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
Table 8-16 Loading Software from the Internal Partition Functional Checks
: Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
Equipment passed all required checks and is ready for
use.
4-3-7 Probe/connector functional checks
4-3-8 B-Mode functional checks
4-3-10 System CFM and PWD functional checks
4-3-20 Basic measurement functional checks
4-2-4 Power shut down

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ Totus and is the answer not in the manual?

GE LOGIQ Totus Specifications

General IconGeneral
BrandGE
ModelLOGIQ Totus
CategoryMedical Equipment
LanguageEnglish

Related product manuals