DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 120 Section 8-7 - Replacing OPIO
8-7-24-2 Removal Procedure
Steps Corresponding Graphic
1. Unscrew 2 screw to remove the USB-C Cable.
Disconnect the USB-A and USB-C connector from
the console.
2. Unscrew 2 screws to remove the User Port Module L
from the console.
3. Unscrew 4 screws to remove the USB Port Fix Brkt
and remove the plastic clamp.
4. Unscrew 2 screws to remove the User Port Module R
from the console.