DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-7 - Replacing OPIO 8 - 121
8-7-24-3 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
5. Unscrew 4 screws to remove the USB Port Fix Brkt
and remove the plastic clamp.
Steps Corresponding Graphic