DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 192 Section 8-11 - Replacing Options
8-11-8-1 Removal Procedure
8-11-8-2 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Step Corresponding graphic
1.
Disconnect the Melody Charger USB Cable from the
console.
2.
Unscrew the NUT and pull the Wireless Probe
Charger from the console.
3.
Unscrew 2 screws to separate Wireless Probe
Charger Set-T and Wireless Probe Charger Holder
Set-T