Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
1 - 26 Section 1-7 - Lockout/Tagout (LOTO)
1-7-2-2 Disconnect Power to the LOGIQ Totus
Table 1-9 Disconnect Power to the LOGIQ Totus
Step Corresponding Graphic
DANGER
Dangerous voltages exist within the LOGIQ Totus.
Hazardous voltages, greater than 30 volts, exist in the LOGIQ Totus.
Always disconnect the power cable from the wall before disconnecting it from the LOGIQ Totus.
1.
To remove the energy supply from the
LOGIQ Totus, do the following:
1. Turn off the breaker on the back of the LOGIQ
Totus.
2. From the wall outlet, disconnect the power
cable.
3. Disconnect the power cable from the LOGIQ
Totus.
4. To prevent unexpected re-energization,
maintain exclusive control over the power cable.
5. Confirm that the ON/OFF button turns OFF.
2.
Disconnect and store any probes attached to the LOGIQ Totus.
3.
Disconnect any external I/O cabling.