Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
10 - 2 Section 10-1 - Overview
10-1-3 Care and maintenance inspections
It has been determined by engineering that your Ultrasound system does not have any high wear
components that fail with use, therefore no care and maintenance inspections are mandatory.
However, some customers’ Quality Assurance Programs may require additional tasks and or
inspections at a different frequency than listed in this manual.
DANGER
There are several places on the backplane, the AC distribution, and DC
distribution that are dangerous. Be sure to disconnect the system power plug
and switch off the main circuit breaker before you remove any parts. Be
cautious whenever power is still on and covers are removed.
DANGER
Do not pull out or insert circuit boards while power is ON.
CAUTION
Practice good ESD prevention. Wear an anti–static strap when handling electronic parts and
even when disconnecting/connecting cables.
CAUTION
DO NOT operate this system unless all board covers and frame panels are securely in place, to
ensure optimal system performance and cooling. When covers are removed, electromagnetic
interference (EMI) may be present.