GE HealthCare CONFIDENTIAL
Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
10 - 6 Section 10-3 - Maintenance Task Schedule
10-3-1 How often should care and maintenance tasks be performed? (continued)
Abbreviations used in the Customer Care Schedule Table 10-1:
D = Daily W = Weekly M = Monthly A = Annually
Table 10-1 Customer Care Schedule
Item Service at Indicated Time D W M A Notes
Air Filter Grid
Clean the air filter grid with vacuum cleaner
from the outside.
•
More frequently depending on
the environment
Remove the filter from the rear and left/right
side cover and clean the hous-ing from the
inside.
•
More frequently depending on
the environment
AC Mains Cable Inspect AC mains cable. • Mobile Unit Check weekly.
Cables and Connectors
Check if all cables are fixed, well seated at
the correct position, and if there is no
mechanical damage visible.
•
Also after corrective
maintenance.
User Interface
Clean alphanumerical keyboard, functional
keys, digital potentiometers.
•
Be careful not to get the cloth
too wet so that moisture does
not enter the entire system.
Display Monitor, Touch
Panel and Probe Holder
Clean the display monitor and Touch panel
surface and probe holder.
•
Be careful not to get the cloth
too wet so that moisture does
not enter the entire system.
Mechanical Parts
Clean and inspect the mechanical function
of wheels, casters, brakes, and swivel locks
as well as side door, foot rest, front and rear
handle, and monitor holder.
Remove dust and coupling gel.
• Mobile Unit Check Daily.
VNAV roll stand
Inspect all fasteners associated with the
mounting system on the VNAV roll stand.
Tighten them as necessary for optimal
operation and safety.
•
DO NOT overtighten any
fasteners.
Control Console
Movement
Check translation/rotation and height
adjustment (elevation).
•
More frequently at Mobile
Units.
Trackball Check
Check proper operation
(Cursor movement X, Y direction).
•
If failure occurs go to trackball
cleaning.
Trackball Cleaning
Remove trackball ring; open the trackball
housing and take out the trackball to clean it
with a soft tissue.
•
Please record it in the systems
setup maintenance report.
Disk Drives
(Data Backup)
Test Image filing (Archive) Import and
Export data capability.
• •*
* Save the image filing data
weekly or at least monthly on
USB depending on the number
of examinations.
Safe Probe Operation
Clean probes and probe cables and check
acoustic lens housing (cracks) and probe
cables. In case of mechanical damage, do
not use them! Danger: Safety risk for
operator and patient.
•* * Or before each use.
Probe Air Bubbles
To detect air bubbles in filling liquid, shake
the probe carefully and check abnormal
noise.
•
Probe Connectors
Remove dust/dirt of all probe connectors.
Clean with vacuum cleaner if dust is visible.
•