EasyManua.ls Logo

GE LOGIQworks - Configure TCP;IP Settings

GE LOGIQworks
420 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
3-76 Section 3-5 - Configuration
3-5-1-4 Configure TCP/IP Settings
Connect your LOGIQworks system to the network.
To achieve communication between Scanner and LOGIQworks the following steps must be completed
successfully.
Configure the TCP/IP settings using the Windows Control Panel:
If not already configured according to the information you provided with the Data Collection Document
(see chapter ) or in case of configuration changes enter the correct data in the fields described in the
figure below.
Please refer to Section 3-5-4-6 "Receive" on page 3-140 for details on how to configure the “receive”
behavior of LOGIQworks through the Connection Service. Also please refer to Section 3-5-10
"Connectivity (Application Level)" on page 3-183 for general Connectivity Settings on the Application
Level.
Figure 3-65 Internet Protocol (TCP/IP)
NOTE
On the scanner the connectivity TCP/IP must be setup with the above TCP/IP information in the area
for a Remote Archive to achieve communication with LOGIQworks.
You need to ask your network administrator for
the appropriate settings.
Type in:
IP address
Subnet mask
Default gateway
NOTE:This example shows fictive numbers!
When you receive your LOGIQworksthe
TCP/IP settings should reflect the
information you provide with the Data
Collection Document.

Table of Contents

Related product manuals