GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5 LOGIQWORKS® BASIC SERVICE MANUAL
Chapter 2 Pre-Installation 2-3
2-2-1-1 Cooling (cont’d)
Table 2-5 LOGIQworks Standard Hardware of System based on Dell Precision 340
2-2-1-2 Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
PART PC CRT1 CRT2 LCD1 LCD2 Modem
BTU/HR 860 495 495 195 195 164
Power (W) 250 135 135 57 57 48 Total BTU/HR
LOGIQworks Models
H49021LA x x x 1520
H49021LB x x x x 2015
H49021LC x x x 1220
H49021LD x x x x 1415