GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
Chapter 4 Functional Checks 4-3
4-3-1 Windows Level Check
4-3-1-1 Check the DICOM Connection
Receiving DICOM images requires preparation of both the receiving and the sending site.
Condition: The LOGIQworks system needs to be connected to the network. See Section 2-3-4 on page
2-14.
Testing the DICOM association:
We recommend testing the association to all devices listed in the destinations and origins table (please
refer to Section 3-5-10-1 "Setting up and configuring origins" on page 3-183 and Section 3-5-10-2
"Setting up and configuring destinations" on page 3-186) to ensure that the remote system and
LOGIQworks have been correctly configured. There is a utility to test the DICOM association: DCMPing,
which can be found in: C:\AMI_61\Bin for LOGIQworks up to 1.3 systems and in directory C:\Program
Files\Centricity\RA600\7.0\Bin for LOGIQworks 2.0 and LOGIQworks 3.0.