GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
3-2 Section 3-1 - Overview
3-1-3 Safety Reminders
WARNING
WARNING
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH
THE UNIT!
WARNING
WARNING
Two people should unpack the unit because of its weight. Two people are required whenever a part
weighing 22 kg (50 lbs.) or more must be lifted.
CAUTION
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate to its
operating environment.
Table 3-2 LOGIQworks Acclimate Time
°C
-40-35-30-25-20-15-10-5 0 5 1015202530354045505560
°F
-40 -31 -22 -13 -4 5 -50 41 32 41 50 59 68 77 86 95 104 113 122 131 140
hrs
201816141210864200000002468
WARNING
WARNING
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use a
three to two prong adapter. This defeats safety grounding.
WARNING
WARNING
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the LOGIQworks and
kept near the unit for quick reference.