EasyManua.ls Logo

GE LOGIQworks - Popup Menu

GE LOGIQworks
420 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5 LOGIQWORKS® BASIC SERVICE MANUAL
Chapter 3 Installation 3-83
3-5-2-3 Logging (cont’d)
Logging Level: Set the level of logging messages. If set to ALL, all messages will be logged. If set
to NORMAL (the typical setting) only error messages will be logged.
Logging Type: Set the types of logging used by the RA 600 application. (Default: File)
- File: If this check box is checked, RA 600 will write general information about its operation to
the RadWorks.log file, which you can view in the Log folder.
- Pipe: Used in combination with the listening tool in the X:\AMI_61\bin folder. Among other
things, it enables you to observe DICOM transmissions in a pipe. Consult the technical
reference manual or contact technical support for additional information.
- Console: If this check box is checked, LOGIQworks will display a window that continuously
shows the actions LOGIQworks is performing. This is useful when troubleshooting.
Log File: Set the name of the general log file. Each Service has its own default log filename. See
the Database Service, Connection Service and Print Service properties for details.
(Default: D:\LOG\AMI_61\RadWorks.LOG; DO NOT CHANGE THE DEFAULT SETTINGS!)
Truncation Size (kB): Set the truncation size of the logging file. The truncation size is the size the
log file will be shortened to when its maximum size has been exceeded.(Default: 16 kB)
Maximum Size (kB): Set the maximum size of the log file.
Set the Maximum Size of the log file to 0 to disable log file truncation. This can be very handy for
troubleshooting. (Default: 64 kB)
Always Flush File after Write: If checked, all logging actions will be written immediately to the log
file. This option will have a serious impact on the performance of LOGIQworks, so it should not
normally be checked. (Default: not checked)
Log Pipe: Name of the pipe where logging information is written.
(Default: \\.\pipe\AMI2\RadWorks\Logging)
3-5-2-4 Popup Menu
To select/deselect popup menu:
1.) Select Configuration > Generic... from the Data Selector menu.
2.) Select the Popup Menu tab from the Generic Properties dialog.

Table of Contents

Related product manuals