DIRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
3-106 Section 3-5 - Configuration
3-5-3-3 Logging (cont’d)
• Logging Level: Set the level of logging messages. If set to ALL, all messages will be logged. If set
to NORMAL (the typical setting) only error messages will be logged.
• Logging Type: Set the types of logging used by the RA 600 application. (Default: File)
- File: If this check box is checked, RA 600 will write general information about its operation to
the RadWorks.log file, which you can view in the Log folder.
- Console: If this check box is checked, LOGIQworks will display a window that continuously
shows the actions LOGIQworks is performing. This is useful when troubleshooting.
• Log File: Set the name of the general log file. Each Service has its own default log filename. See
the Database Service, Connection Service and Print Service properties for details.
(Default: [DATADIR%\log\workstation.log%)]; DO NOT CHANGE THE DEFAULT SETTINGS!)
• Maximum Size (kB): Set the maximum size of the log file.
Set the Maximum Size of the log file to 0 to disable log file truncation. This can be very handy for
troubleshooting. (Default: 64 kB)
• Always Flush File after Write: If checked, all logging actions will be written immediately to the log
file. This option will have a serious impact on the performance of LOGIQworks, so it should not
normally be checked. (Default: not checked)
3-5-3-4 Popup Menu
To select/deselect popup menu:
1.) Select Configuration > User... from the Data Selector menu.
2.) Select the Popup Menu tab from the User Properties dialog.