EasyManua.ls Logo

GE LOGIQworks - Page 174

GE LOGIQworks
420 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
3-108 Section 3-5 - Configuration
3-5-3-5 System Properties (cont’d)
System Measurement: Select the units of measurements you want LOGIQworks to use from the
drop-down list. If you select REGIONAL SETTINGS (see Figure 3-74), LOGIQworks will use the
settings as specified in Regional Settings in the Windows Control Panel (see Section 3-5-1-2 on
page 3-70). (Default: METRIC)
If you wish to overrule Windows' regional settings, you can select either METRIC (Default setting) or US
from the drop-down list. LOGIQworks will then use millimeters or inches irrespective of what is specified
in the Windows Control Panel.
Figure 3-94 Generic Properties - System Settings
System measurement
Administrator only

Table of Contents

Related product manuals