IRECTION 5184024-100, REVISION 5 LOGIQWORKS® BASIC SERVICE MANUAL
Chapter 3 Installation 3-135
3-5-4-3 Logging (cont’d)
Logging Level: Set the level of logging messages. If set to ALL, all messages will be logged. If set
to NORMAL (the typical setting) only error messages will be logged.
Logging Type: Set the types of logging used by the RadWorks application. (Default: File)
- File: If this check box is checked, RadWorks will write general information about its operation
to the RadWorks.log file, which you can view in the Log folder.
- Pipe: Used in combination with the listening tool in the X:\AMI_61\bin folder. Among other
things, it enables you to observe DICOM transmissions in a pipe. Consult the technical
reference manual or contact technical support for additional information.
- Console: If this check box is checked, RadWorks will display a window that continuously
shows the actions RadWorks is performing. This is useful when troubleshooting.
Log File: Set the name of the log file. Each Service has its own default log filename. (Do not
change)
Truncation Size (kB): Set the truncation size of the logging file. The truncation size is the size the
log file will be shortened to when its maximum size has been exceeded. (Default: 800 kB)
Maximum Size (kB): Set the maximum size of the log file. (Default: 1024 kB)
Figure 3-121 Connection Service Properties - Logging (Default Settings)
D:\RA600\LOG