IRECTION 5184024-100, REVISION 5 LOGIQWORKS® BASIC SERVICE MANUAL
Chapter 3 Installation 3-179
3-5-9-2 Logging
To set and view RA600 logging properties:
1.) Select Configuration > Print Service... from the Data Selector menu.
2.) Select the Logging tab from the Print Service Properties page.
Logging Level: Set the level of logging messages. If set to ALL, all messages will be logged. If set
to NORMAL (the typical setting) only error messages will be logged.
Logging Type: Set the types of logging used by the RadWorks application. (Default: File)
- File: If this check box is checked, RadWorks will write general information about its operation
to the RadWorks.log file, which you can view in the Log folder.
- Console: If this check box is checked, RadWorks will display a window that continuously
shows the actions RadWorks is performing. This is useful when troubleshooting.
Log File: Set the name of the log file. Each Service has its own default log filename.
Maximum Size (kB): Set the maximum size of the log file. (Default: 1024 kB)
Set the Maximum Size of the log file to 0 to disable log file truncation. This can be very handy for
troubleshooting.
Figure 3-152 Print Service Properties - Logging (Default settings)