GSP-9300 User Manual
206
3. Press Correction[F2] to apply corrective factors
based on the antenna used, the LISN or the
position of the antenna. The options available
depend on the EMI test frequency selected
previously.
None[F1]
No correction is used. Turn the
Correction function off.
Horizontal[F2]
For radiated emissions tests. It
is a correction set for when the
antenna is in the horizontal
position. This correction set is
configured for use with GW
Instek antennas only.
Vertical[F3]
For radiated emissions tests. It
is a correction set for when the
antenna is in the vertical
position. As with the
Horizontal correction set, this
correction set is configured for
use with GW Instek antennas
only.
Other factor[F4]
Load or create a custom
correction set to use with a
third party antenna.
Cor.factor[F2]
Load or create a custom
correction set for use with a
LISN device.
4. If Other factor[F4] or Cor.factor[F2] was chosen
above, select a correction set or edit an existing
correction set and then set Correction[F2] to On.
Press Return[F7] to return to the previous
menu.
See page 54 for details on creating or editing
correction sets.