ADVANCED OPERATION
207
5. Press Recall Limit[F3] to add EN55022A/B or
FCC A/B limit lines to the display. The limit
line that is produced is matched to the selected
test frequency range.
None
No limit line
EN5502A
Euro commercial standard
EN5502B
Euro residential standard
FCC A
American non-residential standard
FCC B
American residential standard
6. Select Average[F1] or Quasi-peak[F2] for the
detector for the chosen standard. The
availability of these settings depends on the
EMI test frequency selected previously.
Detector:
Average, Quasi-peak
7. When Recall Limit is turned on, pass/fail
testing will be performed on each sweep based
on the standard selected.
Pass:
, with green grid border.
Fail:
, with red grid border.
The alarm icon is shown at the bottom of
the display whenever Recall Limit is turned
on.