ADVANCED OPERATION
213
With the DUT turned on, touch the EMI-E probe
anywhere on one of the PCB traces, PCB pins, I/O
cable pins or one of the other exposed conductors.
When using an EMI-E probe, please make use of a
DC block or other limiter to protect the RF input of
the spectrum analyzer.
> EMI E Probe[F3]> and choose a
frequency band to test at.
2. Next select the type of e-field source that will be
used with the E-field probe.
For use on PCB traces/pins
For use on I/O cabling/pins.
3. Press Correction[F1] and choose the length of
the PCB trace/I/O cable. This will help the
software to estimate the radiated emissions that
would emanate from those points.
For traces of approx. 20cm
For traces of approx. 15cm
For traces of approx. 10cm