5.2 Connecting to the Lines to be Measured
84
NU1U2U3
I1 I2 I3
The arrows point toward the load.
Source Side
Load Side
R
S
T
U
2
U
1
U
3
R, S, T: Line
Three-Phase 3-Wire (3P3W2M)
NU1U2U3
I1 I2 I3
The arrows point toward the load.
Source Side
Load Side
R
S
T
U
2
U
1
U
3
R, S, T: Line
Three-Phase 3-Wire (3P3W3M)