81
────────────────────────────────────────────────────
7.4 Message Reference
────────────────────────────────────────────────────
Synta
x
HEADer data
dat
a
ON/OFF (Character data)
Exampl
e
Transmission
HEAD OFF
No header is affixed to the
response message.
Synta
x
:HEADer?
Respons
e
synta
x
(If headers are ON)
:HEADER ON
(If headers are OFF)
OFF
Exampl
e
(If headers are ON)
Transmission
:HEAD?
Response
:HEADER ON
(If headers are OFF)
Transmission
:HEAD?
Response
OFF
Synta
x
:HOLD data
dat
a
ON/OFF (Character data)
Exampl
e
Transmission
:HOLD ON
The measured value can be held.
F
unctio
n
This command turns response
message header on or off.
F
unctio
n
This command returns the response
message header setting in ON/OFF
character data (data) format.
F
unctio
n
This command turns measurement
hold on or off.
Note
s
When the measurement mode is
changed in the hold status, the
displayed voltage may not be output
to the RS-232C and GP-IB, or
irrelevant values may be displayed
on the screen. After the
measurement mode is changed,
measure the waveforms again.
:HEADer
Enable and disable headers.
:HEADer?
Queries the headers enablement.
:HOLD
Sets the measurement hold.

Table of Contents

Related product manuals