47
────────────────────────────────────────────────────
6.3 Measurement by External Control Terminal and External Output Terminal
────────────────────────────────────────────────────
.3.3 Timing Chart
〜
Description
Time (approx.)
MAX TYP MIN
t
1
1ms − −
t
2
6ms − 4ms
t
3
Sampling rate
FAST 20 ms(50 Hz) / 16.7 ms(60 Hz)
MEDIUM 160 ms(50 Hz) / 133 ms(60 Hz)
SLOW 640 ms(50 Hz) / 533 ms(60 Hz)
t
4
Length of time from measurement trigger to output of EOC
1ms − −
t
5
Measurement-trigger pulse width
− − 1ms
t
6
Measurement trigger interval
− − 1 sampling
t
7
Length of time from
MANU
to output of comparator
1ms − −
t
8
MANU
pulse width
− − 1ms
t
9
Length of time from connection of the 3560 to the sample to
setting of NG to Lo
− 5ms −
t
10
Length of time from disconnection of the 3560 from the
sample to setting of NG to
1ms − −
t
11
0ADJ
,
HOLD
pulse width
− −
15 ms
t
12
PRINT
pulse width
− − 1ms
t
13
Length of time from input to change in comparator number
2s − −
t
14
Input retention time
− − 2s