EasyManua.ls Logo

Hioki 3560 - Timing Chart

Hioki 3560
156 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
47
────────────────────────────────────────────────────
6.3 Measurement by External Control Terminal and External Output Terminal
────────────────────────────────────────────────────
6
.3.3 Timing Chart
H
H
L
H
HOLD
L
OADJ
PRINT
Hold
Chuck
TRIG
MANU
Output
Input
NG
EOC
Hi,IN,Lo
Hi,IN,Lo
Display
1
0
State 
Comparator table
Input 
COMP0
〜
COMP4
t
11
t
12
t
11
t
14
t
13
Free run
Hold
No connection
Connection
No connection
Connection
Effective
Measurement value
t
4
Abnormal measurements
Measurement value
Effective
Effective
t
9
t
10
t
2
t
3
t
1
t
7
t
5
t
7
t
4
t
8
t
6
1
2
1
3
2
3
0
State
H
L
t
11
Effective
H
L
1
0
1
0
L
Description
Time (approx.)
MAX TYP MIN
t
1
1ms − −
t
2
6ms − 4ms
t
3
Sampling rate
FAST 20 ms(50 Hz) / 16.7 ms(60 Hz)
MEDIUM 160 ms(50 Hz) / 133 ms(60 Hz)
SLOW 640 ms(50 Hz) / 533 ms(60 Hz)
t
4
Length of time from measurement trigger to output of EOC
1ms − −
t
5
Measurement-trigger pulse width
− − 1ms
t
6
Measurement trigger interval
− − 1 sampling
t
7
Length of time from
MANU
―――――
to output of comparator
1ms − −
t
8
MANU
―――――
pulse width
− − 1ms
t
9
Length of time from connection of the 3560 to the sample to
setting of NG to Lo
− 5ms −
t
10
Length of time from disconnection of the 3560 from the
sample to setting of NG to
Hi
1ms − −
t
11
0ADJ
―――――
,
HOLD
―――――
pulse width
− −
15 ms
t
12
PRINT
―――――
pulse width
− − 1ms
t
13
Length of time from input to change in comparator number
2s − −
t
14
Input retention time
− − 2s

Table of Contents