72
────────────────────────────────────────────────────
7.4 Message Reference
────────────────────────────────────────────────────
Synta
x
:ADJust?
Respons
e
synta
x
(If headers are ON)
:ADJUST data
(If headers are OFF)
data
Exampl
e
(If headers are ON)
Transmission
:ADJ?
Response
:ADJUST 0
(If headers are OFF)
Transmission
:ADJ?
Response
0
The zero-adjust executes
successfully.
Synta
x
:AUTOrange data
dat
a
ON/OFF (Character data)
Exampl
e
Transmission
:AUTO ON
The measurement is performed in
auto-range mode.
F
unctio
n
This command executes a zero-
adjust and returns a result in 0/1
NR1 numeric data (data) format.
If "data =0,
" the zero-adjust executes correctly.
If "data =1,
" the zero-adjust does not execute
correctly.
Note
s
The zero-adjust cannot be performed
in the following cases:
A zero-adjust is attempted when the
following parameters exceed the
specified limits:
Resistance function
240 dgt. (FAST mode)
2400 dgt. (MEDIUM/SLOW mode
)
Voltage function
3400 dgt. (full sampling)
Abnormal measurements
F
unctio
n
This command turns auto range on
or off.
Note
s
Setting only resistance range or onl
y
voltage range is not permitted.
7
.4.2 Specific Command Messages
:ADJust?
Executes zero-adjust and queries the result.
:AUTOrange
Sets the auto range.