5 - 47
Code Error Message Countermeasure
1574 The power supply of +24V was
shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (on the IO-DRV00 board, F5, F6, F7, F8, or
F9, or more than one fuse) was found.
Click the [OK] button to close the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1575 The power supply of +18V for
C2 was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F14 on the COLHVC00 board) was found.
Click the [OK] button to close the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1576 The power supply of +18V for
C1 was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F13 on the COLHVC00 board) was found.
Click the [OK] button to close the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1577 The power supply of +18V for
OBJ was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F12 on the COLHVC00 board) was found.
Click the [OK] button to close the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1578 The power supply of -15V for
STG was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F11 on the COLHVC00 board) was found.
Click the [OK] button to close the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1579 The power supply of +15V for
STG was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F10 on the COLHVC00 board) was found.
Click the [OK] button to close the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.