5 - 48
Code Error Message Countermeasure
1580 The power supply of -15V for
DEF was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F9 on the COLHVC00 board) was found.
Click the [OK] button to close the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1581 The power supply of +15V for
DEF was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F8 on the COLHVC00 board) was found.
Click the [OK] button to close the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1582 The power supply of +12V for
HVT was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F7 on the COLHVC00 board) was found.
Click the [OK] button to close the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1583 The power supply of -15V for
HVC was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F6 on the COLHVC00 board) was found.
Click the [OK] button to close the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1584 The power supply of +15V for
HVC was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F5 on the COLHVC00 board) was found.
Click the [OK] button to close the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.