5 - 49
Code Error Message Countermeasure
1600 The amount of the image shift
movement exceeds 50 µm.
Reset the image shift, and
move the stage.
The message appears when the image shift
function is not enabled and a specified position
can not be moved to the center of the window.
Click the [OK] button to close the message.
The range over which the FOV can be moved by
Image shift is limited; if the FOV cannot be
moved further, click the [Reset] button once, and
use the XY knob to move the FOV to conduct an
observation.
1611 Time has run out.
“Stage initialized” cannot be
executed.
Start up the main unit and
application again. Contact
service if the failure occurs
repeatedly.
This message appears if a time-out (300 s) has
occurred during the initialization of the motor
stage.
Click the [OK] button to close the message.
1690 The instruction manual does
not exist.
Install the instruction manual.
This message appears when the TM3030Plus
electronic manual (in PDF format) is not installed
on the PC and so it cannot be displayed.
Install the TM3030Plus electronic manual from
the supplied TM3030Plus SETUP CD-ROM.
1691 The instruction manual cannot
be displayed.
Confirm if the PDF file browser
has been installed.
This message appears when the PDF file read
software (Adobe Acrobat Reader) is not installed
on the PC and so the electronic instruction
manual cannot be displayed.
Install the PDF file read software (Adobe
Acrobat Reader) from the supplied TM3030Plus
SETUP CD-ROM.
1706 Observation screen size is
changed.
Do you save this setting?
This confirmation message appears when not
the [SET] button but the [Close] button is clicked
after changing the observation screen size.
Click the [OK] button to change the size, and
close the Setup window. Or, click the [Cancel]
button to close the window without changing the
size.