N9030B PXA Signal Analyzer Service Guide 103
Instrument Messages
Event Messages
–221 Settings conflict; Gate is
not available when Marker
Count on
The gate function cannot be used while you have marker
count turned on.
–221 Settings conflict; Gate
Length is not settable in
FFT sweeps
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
–221 Settings conflict; Gate
Length is not settable in
FFT sweeps
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
–221 Settings conflict; Gate
Method is not compatible
with current Sweep Type
setting
If the Gate is On and you have the FFT Sweep Type manually
selected, then the Gate Method cannot be selected.
–221 Settings conflict; Gate not
available with external
Tracking Source
The Gate functions are unavailable when Source Mode is
Tracking with an external source. This is because the Gate
circuitry is used to sync the external source.
–221 Settings conflict; Gate not
available with Tracking
Generator
If the Source Type is Tracking Generator, the Gate circuitry is
used for TG sync and is not available for gating
–221 Settings conflict; Gated FFT
is not available while Sweep
Type is set to Swept
The gated FFT function is not available if you have selected
the swept type of sweep. You must be in the FFT sweep type.
–221 Settings conflict; Gated LO
is not available while Sweep
Type is set to FFT
The FFT sweep type moves the LO frequency in steps. So the
gated LO function is not available if you have selected FFT
sweep.
–221 Settings conflict; Gated
Video is not available while
Sweep Type is set to FFT
The gated video function is not available if you have selected
the FFT sweep type.
–221 Settings conflict; Incorrect
RBW for demod. Change RBW
–221 Settings conflict; Ind I/Q
is not available for this
measurement
The Independent I and Q setting is not available for the
current measurement. Only some measurements (initially,
only VXA) support this setting.
–221 Settings conflict; Invalid
trace number
The subopcode used to specify the trace number is invalid
for this measurement or query
–221 Settings conflict; Knob is
not available to modify this
function
You should select a specific value for this function. So
scrolling through values with the knob is not allowed.
–221 Settings conflict; LO Phase
Noise Adj not available
For instruments without the Dual-Loop LO, this feature is not
available
Err# Message Verbose/Correction Information

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals