104 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–221 Settings conflict; Log Scale
Type is not available with
Demod View
The logarithmic x-axis scales are not available when you
have the demod view turned on.
–221 Settings conflict; Log Scale
Type is only available in
swept measurement
Logarithmic scaling can be used when making a swept SA
measurement. It is not available in the SA measurement
when you are using FFT sweeps.
–221 Settings conflict; Marker 1
Trace Update=off turns off
Signal Track
Signal Track not available unless the trace containing Marker
1 is updating
–221 Settings conflict; Marker
cannot be relative to itself
A marker must be set relative to another marker, not to
itself.
–221 Settings conflict; Marker
Count is not available when
Gate on
The marker count function cannot be used while you have
gating turned on.
–221 Settings conflict; Marker
Function is not available
for a Fixed marker
If a Marker is a Fixed type marker, the marker's value does
not change from when it first became fixed. You cannot turn
on or change a Marker Function because there is no ongoing
measurement data to use for the marker function
calculation.
–221 Settings conflict; Marker
type must be delta
MkrΔ->Span and MkrΔ->CF require that the selected
marker be a delta marker.
–221 Settings conflict; Marker->
function is not available in
zero span
Most of the “Marker To” functions are not available if you are
in zero span (span = 0 Hz, or time domain). So you cannot
send the commands for these functions.
–221 Settings conflict; Mask
unavailable for current
Span. Increase to display
mask.
The current span setting is either narrower than the mask
width or so wide that there are too few display points to
allow the mask to be drawn. Increase or decrease the span
to display the mask.
–221 Settings conflict; Meas Type
was changed to Examine for
Exp Avg Mode.
Average Mode has been changed to Exponential. Full Meas
Type is not available for Exponential Average Mode therefore
Meas Type has been changed to Examine.
–221 Settings conflict; Meas Type
was changed to Full for
Repeat Avg Mode
Average Mode has been changed to Repeat. Examine Meas
Type is not available for Repeat Average Mode therefore
Meas Type has been changed to Full.
–221 Settings Conflict;
MinPts/RBW limit not met
–221 Settings conflict; Mkr -> CF
is not available when the
x-axis is time domain
The marker to center frequency functionality does not work
when the x-axis is in the time domain.
–221 Settings conflict; MS gain
is not available in this
Mode
Mobile Station gain correction is not available in some
Modes, or in some measurements (for example, the SA
measurement).
Err# Message Verbose/Correction Information