N9030B PXA Signal Analyzer Service Guide 113
Instrument Messages
Event Messages
–222 Data out of range; Two
entries already exist at
this x-axis value.
When entering values for limit lines, you cannot have more than
two y-axis (amplitude) values entered for a specific x-axis
(frequency) value.
–223 Too much data A data element (of block, expression, array type, or string type)
had more data then allowed by the command, or by the available
memory.
–223 Too much data; 200 spurs
found. Additional spurs
ignored.
There are too many spurs for the table (the limit is 200), and any
additional spurs that are found will be ignored.
–224 Illegal parameter value An exact data value (from a list of the allowed values) was required
- but not found. See the feature description for information about
the expected parameter values.
–224 Illegal parameter value;
<Value> invalid.
Fractional values are not
allowed.
The seconds parameter of an LXI time may not contain a fractional
portion. For example 123456789.0 is valid while 123456789.1 is
not.
–224 Illegal parameter value;
<value> out of range.
The value does not fall in the valid range
–224 Illegal parameter value;
Exceeding the max list
length
The list parameters have a maximum allowed length. You are
trying to set a length longer than the maximum.
–224 Illegal parameter value;
Gated FFT is not available
while Sweep Type is set to
Swept
The gated FFT function is not available if you have selected the
swept type of sweep. You must be in the FFT sweep type.
–224 Illegal parameter value;
Gated LO is not available
while Sweep Type is set to
FFT
The FFT sweep type moves the LO frequency in steps. So the
gated LO function is not available if you have selected FFT sweep.
–224 Illegal parameter value;
Gated Video is not
available while Sweep Type
is set to FFT
The gated video function is not available if you have selected the
FFT sweep type.
–224 Illegal parameter value;
Illegal identifier
<identifier>. This value
may already be in use.
The value for the LXI LAN identifier parameter must be unique (i.e.
LAN0 and LAN7 must have different identifier strings).
–224 Illegal parameter value;
Index out of range
When querying the LXI Event Log or the Servo Log, an index may
be used to look at a specific entry. This error occurs if the index
provided does not point to a valid entry.
Err# Message Verbose/Correction Information