N9030B PXA Signal Analyzer Service Guide 117
Instrument Messages
Event Messages
–250 Mass storage error; Pk
Table must be on to save
Pk Table as Meas Results
You have to have a Peak Table on the screen before you can save
it. Turn on the Peak Table and try again.
–250 Mass storage error; Read
fault
The system cannot read from the specified device.
–250 Mass storage error;
Register <number> empty
Attempt to recall a register with nothing in it
–250 Mass storage error;
Sharing violation
The process cannot access the file because it is being used by
another process.
–250 Mass storage error;
Spectrogram must be on to
save as Meas Results
You have to have a Spectrogram on the screen before you can
save it. Turn on the Spectrogram and try again.
–250 Mass storage error; Too
many open files
The system cannot open the file.
–250 Mass storage error; Write
fault
The system cannot write to the specified device.
–252 Missing media A legal command or query could not be executed because
missing media.
–253 Corrupt media A removable media was found to be bad or incorrectly formatted.
Any existing data on the media may have been lost.
–254 Media full A legal command/query could not be executed because the
media was full
–255 Directory full A legal command or query could not be executed because media
directory was full.
–256 File name not found; A legal command or query could not be executed because the file
name was not found in the specified location.
–257 File name error; A legal command or query could not be executed because there
was an error with the file name on the device media. For example,
maybe you tried to copy to a duplicate file name.
–257 File name error; Allowable
extension is .csv
You are using the wrong type of file extension for the current
data/file type.
–257 File name error; Allowable
extension is .png
You are using the wrong type of file extension for the current
data/file type.
–257 File name error; Allowable
extension is .state
You are using the wrong type of file extension for the current
data/file type.
–257 File name error; Invalid
file name
The filename, directory name, or volume label syntax is incorrect.
–257 File name error; name too
long
Err# Message Verbose/Correction Information

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals