116 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–250 Mass storage error; Can
only load an Antenna Unit
into Correction 1
The only Correction register that supports Antenna Units is
number 1. You have attempted to load an Ampcor file which
contains antenna units into another register
–250 Mass storage error; Cannot
make
The directory or file cannot be created.
–250 Mass storage error;
Different Antenna Unit
already in use
Attempt to import Corrections file with Antenna Unit that differs
from an in-use correction.
–250 Mass storage error;
Directory not found
The system cannot find the path specified.
–250 Mass storage error; Failed
to Load trace. Bad file
format.
The load trace operation could not be completed, as the input file
was not in the expected format. You can only load traces that were
previously saved using the 'Save Trace' feature.
–250 Mass storage error; File
<filename> wrong type
Attempt to import a data file that is not the proper type for this
operation.
–250 Mass storage error; File
<filename> and instrument
version mismatch
While opening a file, there was a mismatch between file version or
model number with instrument version or model number. The
import still tried to load as much as possible, but you should
check it closely.
–250 Mass storage error; File
contains incorrect data
for this operation
There is a mismatch between the file data type of the file specified
and the destination indicated. For example, a correction set
cannot be loaded/imported into a limit line.
–250 Mass storage error; File
empty
Cannot save trace because it contains no data. Check that the
trace is turned on and contains some valid data.
–250 Mass storage error;
Invalid register number
for *SAV or *RCL Mass
Storage error
You have used the *SAV command to save a state to a non
existent state register.
Or
You have used the *RCL command to recall a state register that
wasnt previously saved with the *SAV command.
–250 Mass storage error; Lock
violation
The process cannot access the file because another process has
locked a portion of the file.
–250 Mass storage error; Mkr
Table must be on to save
Mkr Table as Meas Results
You have to have a Marker Table on the screen before you can
save it. Turn on the Marker Table and try again.
–250 Mass storage error; No
file names available
Attempt to use the auto file name generation when all 10,000 file
names are taken.
–250 Mass storage error; Open
failed
The system cannot open the device or file specified. This could be
because the storage media is full, or possibly due to a filename
error. If using an external storage devise, check that the devise is
properly formatted.