N9030B PXA Signal Analyzer Service Guide 115
Instrument Messages
Event Messages
–230 Data corrupt or stale;
Unable to load state from
file
There is something wrong with the state data in the desired file.
Maybe the file is corrupt, or it is from an instrument/version that is
not recognized by the current instrument.
–231 Data questionable Indicates that the measurement accuracy is suspect
–232 Invalid format A data element was found but it could not be used because the
data format or the data structure was not correct.
–232 Invalid format; Map
information not loaded
Instrument failed to load the burst mapping information from the
selected file.
–233 Invalid version A legal data element was found but could not be used because
the version of the data is incorrect. For example, state data
changes as new instrument features are added, so old state files
may not work in an instrument with a newer version of software.
–240 Hardware error A legal program command or query could not be executed
because of a hardware error. The exact problem cannot be
specifically identified.
–240 Hardware error; See
details in Windows Event
Log under SA
The internal data acquisition system detected a problem at
startup and logged the details in the Windows Event Log.
–241 Hardware missing The operation could not be performed because of missing
hardware; perhaps the optional hardware is not installed.
–241 Hardware missing; Internal
preamp not available at
all frequency points
The Internal Preamp is currently turned on, but the measurement
is being performed completely or partially outside the range of the
preamp. It is recommended that the user turns preamp off to
ensure consistent results across the entire measurement.
–241 Hardware missing; not
available for this model
number
The hardware required is not part of this model
–241 Hardware missing; Option
not installed
The optional hardware is not installed.
–250 Mass storage error; A problem was found with the mass storage device (memory, disk
drive, etc.). The exact problem cannot be specifically identified.
–250 Mass storage error; Access
denied
Access is denied.
–250 Mass storage error; Bad
path name
The specified path is invalid.
–250 Mass storage error; Can
only import single trace
.csv files
Trace files containing multiple traces can not be imported.
However, if you need to recall multiple traces you can use the
Save and Recall functions rather than the Import and Export
functions.