N9030B PXA Signal Analyzer Service Guide 37
Overview
Service Equipment You Will Need
3 dB
Fixed Attenuator
Frequency: 50 MHz
Loss: 3 dB (nominal)
VSWR: at 50 MHz: ≤ 1.2:1
Connector: Type-N (m, f)
8491A
(Option 003)
8491B
(Option 003)
P
6 dB
Fixed Attenuator
Frequency: 50 MHz
Loss: 6 dB (nominal)
VSWR: at 50 MHz: ≤ 1.05:1
Connector: Type-N (m, f)
8491A
(Option 006, H33
i
)
8491B
(Option 006,
H33
i
)
P
6 dB
Fixed Attenuator
(2 required)
Frequency: 50 MHz to 3.6 GHz
Loss: 6 dB (nominal)
VSWR: at 50 MHz: ≤ 1.10:1
Connector: 3.5 mm (m, f)
8493C
(Option 006)
A
10 dB
Fixed Attenuator
Frequency: 50 MHz
Loss: 10 dB (nominal)
VSWR: at 50 MHz: ≤ 1.05:1
Connector: Type-N (m, f)
8491A
(Option 010, H33
i
)
8491B
(Option 010,
H33
i
)
P
10 dB
Fixed Attenuator
Frequency: 50 MHz to 19.5 GHz
Loss: 10 dB (nominal)
VSWR: at 321.4 MHz to 8 GHz: ≤ 1.1:1
8 GHz to 12.4 GHz: < 1.15:1
12.4 GHz to 19.5 GHz: < 1.25:1
Connector: 3.5 mm (m, f)
8493C
(Option 010)
A, P
20 dB
Fixed Attenuator
Frequency: 10 MHz
Loss: 20 dB (nominal)
VSWR: 10 MHz: ≤ 1.20:1
Connector: Type-N (m, f)
8491A
(Option 020)
8491B
(Option 020)
A, P
30 dB
Fixed Attenuator
Frequency: 50 MHz
Loss: 30 dB ±.0.05 dB
VSWR: at 50 MHz: ≤ 1.05:1
(For use with Low Power Sensors)
11708A A, P
Terminations
Type-N (m)
(for Option 503, 508, 513)
Frequency: 10 kHz to 18 GHz
VSWR: ≤ 1.05:1
909A
(Option 012)
P, T
2.4 mm (f)
(for Options > 526)
50 W
Frequency: 10 kHz to 50 GHz
85138B P
3.5 mm (f)
(for Option 526)
50 W
Frequency: 10 kHz to 50 GHz
909D
(Option 526)
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Keysight Model
Number
a
or Vendor
Model Number
Alternative
Keysight Model
Number
a
Use
b