N9030B PXA Signal Analyzer Service Guide 39
Overview
Service Equipment You Will Need
Cables
3.5 mm (m) to 3.5 mm (m)
(2 required)
Frequency: DC to 26.5 GHz
Length: ≤ 92 cm (36 in)
Insertion Loss: ~2 dB
VSWR: DC to 18 GHz: ≤ 1.25:1
18 GHz to 26.5 GHz: ≤ 1.35:1
11500E 8120-4921 A, P
2.4 mm (f) to 2.4 mm (m)
(for Options > 526)
Frequency: DC to 50 GHz
Length: ≤ 24.9 mm (9.8 in)
Insertion Loss:
≤ 26 GHz, ~4 dB
≤ 40 GHz, ~5 dB
≤ 50 GHz, ~6 dB
VSWR:
≤ 26.5 GHz: ≤ 1.30:1
≤ 40 GHz: ≤ 1.40:1
≤ 50 GHz: ≤1.55:1
8120-6164 A, P
Type-N
(2 required)
Frequency: 10 MHz to 8 GHz
Precision Type-N (m), both ends
62 cm (24 in.)
VSWR: ≤ 18 GHz: 1.4:1
Insertion Loss: 1.5 dB
11500C A, P, T
BNC
(3 required)
Frequency: DC to 10 MHz
50 Ω Coax BNC (m), both ends
120 cm (48 in.)
10503A A, P, T
Filters
50 MHz Low Pass Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Insertion Loss: ~1 dB
VSWR: ≤ 1.5:1
BNC (m) to BNC (f)
Telonic Berkeley
TLA 50-5AB2
0955-0306 P,T
300 MHz Low Pass
(2 Required)
Cutoff Frequency: 300 MHz
Rejection at > 435 MHz: > 45 dB
VSWR: ≤ 1.5:1
BNC (m, f)
Telonic Berkeley
TLP 300-4AB4
0955-0455 P
1.8 GHz Low Pass
(2 Required)
Cutoff frequency: 1.8 GHz
Rejection at 2.1 GHz to 6 GHz: > 45 dB
Insertion Loss at 1.8 GHz: < 1 dB
VSWR: ≤ 1.50:1, DC to 1.8 GHz
Connector: SMA (f) to SMA (f)
RLC L-1636 0955-0491 P
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Keysight Model
Number
a
or Vendor
Model Number
Alternative
Keysight Model
Number
a
Use
b