N9030B PXA Signal Analyzer Service Guide 439
Assembly Replacement Procedures
RF Area (Options 503, 508, 513, 526)
3. Refer to Figure 16-7. Remove the RF bracket (1) by removing the sixteen
screws (2) (0515-0372) using the T-10 driver. If the instrument has Option
RTS installed, there will be an additional screw (3) to be removed.
Figure 16-7 RF Bracket Removal