584 N9030B PXA Signal Analyzer Service Guide
Functional Tests
Second Harmonic Distortion (SHD)
9. Press Peak Search. Enter the displayed value under the Measured Second
Harmonic Distortion (dBc) heading in Table 18-4.
Table 18-4 Second Harmonic Distortion Results
Applied Frequency
(MHz)
Measured Second Harmonic Distortion
(dBc)
Mixer Level
(dBm)
Specification
(dBc)
40 –15 –60