N9030B PXA Signal Analyzer Service Guide 605
Functional Tests
BBIQ Input Frequency Response (Option BBA only)
14.Press Peak Search on the X-Series analyzer. Record the marker amplitude
reading in Table 18-10 for the I Input at 250 kHz.
15.Set the RF Signal Generator to each of the remaining frequencies listed in
Table 18-10, up to the maximum IQ bandwidth of the X-Series analyzer. At
each frequency, do the following:
a. Press Peak Search
b. Record the marker amplitude reading in Table 18-10 for the I Input
at the current RF Signal Generator frequency
Table 18-10 BBIQ Input Frequency Response, I and I Inputs
RF Signal
Generator
Frequency
I Input I Input Test Limits
Marker
Amplitude
Frequency
Response
Marker
Amplitude
Frequency
Response
250 kHz 0.0 dB 0.0 dB ±1.75 dB
2 MHz ±1.75 dB
5 MHz ±1.75 dB
7.5 MHz ±1.75 dB
9.9 MHz ±1.75 dB
Stop here if Maximum IQ Bandwidth is 10 MHz
12.5 MHz ±1.75 dB
15 MHz ±1.75 dB
17.5 MHz ±1.75 dB
20 MHz ±1.75 dB
22.5 MHz ±1.75 dB
24.9 MHz ±1.75 dB
Stop here if Maximum IQ Bandwidth is 25 MHz
30 MHz ±1.75 dB
35 MHz ±1.75 dB
39.9 MHz ±1.75 dB