Keysight EXG and MXG X-Series Signal Generators User’s Guide 91
Optimizing Performance for All Models
Using the Dual Power Meter Display (with USB U2000A Series Power Sensors)
13.On the signal generator: Press Channel B to On and then back to Off again. This initializes the
signal generator to the external power meter.
14.Press Return > Zero Sensor
A diagnostic dialog box is displayed each time an external power meter is being used and the
Zero Sensor or Calibrate Sensor softkey is pressed (refer to Figure 5-10 on page 91).
Verify the power sensor is connected to the 50 MHz reference of the power meter.
Figure 5-10 Diagnostic Dialog Box for Channel B
15.Press Done
The Running Calibration(s) bar is displayed: Zeroing Sensor Please wait....
16.After Running Calibration(s) bar disappears: Press Calibrate Sensor
Diagnostic Dialog box is displayed that prompts for verifying the connection of the power
sensor to the power meter 50 MHz reference (refer to Figure 5-11 on page 91).
Figure 5-11 Diagnostic Dialog Box for Calibration
For details on
each key, use key
help as described
on page56.
For details on
each key, use key
help as described
on page56.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals