106 Keysight B2980B User’s Guide, Edition 1
How to Perform Measurement
Resistance Measurement
Figure 3-14 Grounded Device Measurement
The connection shown in Figure 3-14 applies the voltage to the Common terminal
from the Voltage Source. The voltage up to 500 V can be applied to the Common
terminal.
Voltage Source Low terminal is internally connected to or disconnected from the
circuit common as shown in Figure 3-14. See “Low Terminal State” on page 110
for more information.
Input
Triaxial cable
Triaxial bulkhead connector
RX
Low
Common is connected to Voltage Source High by using Banana-to-banana cable.
Shield
High voltage test lead
Guard shield
Low
RX
Input
A
Common
VS
Voltage Source Low Terminal State:
FLOATING
Shield
Guard
shield