EasyManuals Logo

Keysight B2980B Series User Manual

Keysight B2980B Series
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
106 Keysight B2980B User’s Guide, Edition 1
How to Perform Measurement
Resistance Measurement
Figure 3-14 Grounded Device Measurement
The connection shown in Figure 3-14 applies the voltage to the Common terminal
from the Voltage Source. The voltage up to 500 V can be applied to the Common
terminal.
Voltage Source Low terminal is internally connected to or disconnected from the
circuit common as shown in Figure 3-14. See “Low Terminal State” on page 110
for more information.
Input
Triaxial cable
Triaxial bulkhead connector
RX
Low
Common is connected to Voltage Source High by using Banana-to-banana cable.
Shield
High voltage test lead
Guard shield
Low
RX
Input
A
Common
VS
Voltage Source Low Terminal State:
FLOATING
Shield
Guard
shield

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight B2980B Series and is the answer not in the manual?

Keysight B2980B Series Specifications

General IconGeneral
BrandKeysight
ModelB2980B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals