Mastersizer 3000 Page 6-1
6
Sample preparation
guidance
Introduction
Preparation of the sample before it is added to the system can be critical. Over half
the problems encountered when measuring a sample are caused by poor sample
preparation. If the sample is sticking together, dissolving or floating on the surface,
or if it is not a representative sample, the result will be incorrect.
There are many techniques available to ensure that the sample is prepared success-
fully. Once a suitable dispersion technique is found for the sample, standardise the
procedure (using an SOP) so that comparisons can be made between samples.
This chapter gives information on:
Sample preparation flow chart – summarises the process.
Representative sampling.
Considerations for dry samples.
Considerations for wet samples.
Symptoms of poor sample preparation.

Table of Contents

Other manuals for Malvern Mastersizer 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Malvern Mastersizer 3000 and is the answer not in the manual?

Malvern Mastersizer 3000 Specifications

General IconGeneral
Measurement principleLaser diffraction
Blue laser wavelength470 nm
Sample dispersion optionsWet and dry dispersion
Data acquisition rate10 kHz
Particle size range0.01 to 3500 microns
Laser typeHe-Ne and solid state
AccuracyBetter than 1%
RepeatabilityBetter than 0.5%
TechnologyMie and Fraunhofer theories
Sample dispersion unitsHydro MV, Hydro LV, Aero S
Optical modelMie theory
Power requirements100-240 V, 50/60 Hz
Detector arrangementLog-spaced array

Related product manuals