Making measurements Chapter 4
Mastersizer 3000 Page 4-45
Data processing - Result
The results settings allow you to specify the displayed size range and the type of
distribution that the analysis has used in the measurement.
Section/option Description
Result range Reducing the measurement range is not usually a recom-
mend option, especially when conform to good laboratory
practice. This facility is useful though if a sample includes
two or more distinct size populations which may need to be
monitored individually.
Adjusting the low and high size range will alter the fit
applied to the fundamental result, where the intermediate
values are then read from the curve allowing interpolation of
percentile points. This adjustment will not alter the analysis
bands that are used in the measurement.

Table of Contents

Other manuals for Malvern Mastersizer 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Malvern Mastersizer 3000 and is the answer not in the manual?

Malvern Mastersizer 3000 Specifications

General IconGeneral
Measurement principleLaser diffraction
Blue laser wavelength470 nm
Sample dispersion optionsWet and dry dispersion
Data acquisition rate10 kHz
Particle size range0.01 to 3500 microns
Laser typeHe-Ne and solid state
AccuracyBetter than 1%
RepeatabilityBetter than 0.5%
TechnologyMie and Fraunhofer theories
Sample dispersion unitsHydro MV, Hydro LV, Aero S
Optical modelMie theory
Power requirements100-240 V, 50/60 Hz
Detector arrangementLog-spaced array

Related product manuals