Chapter 4 Making measurements
Page 4-38 MAN 0474
Sample Dispersion - Cleaning
The Cleaning options allow you to specify both a clean sequence and whether to
use ultrasound as part of the sequence. This ensures that all traces of the sample
just measured are removed so that no contamination of the next sample occurs.
Cleaning following a measurement is essential to ensure that background noise,
consisting of particles agglomerating within the system and forming accumula-
tions, is minimized.