Page #134 background imageLoading...
Page #134 background image
Chapter 6 Sample preparation guidance
Page 6-2 MAN 0474
6
Sample preparation flow chart
This flow diagram shows the route taken to prepare an unknown sample:
ill 7825
Using liquid-borne particle suspensions, such as latex and emulsions, is usually
straight forward. Use the suspension phase as the dispersant, mix the sample well
and add directly to the Mastersizer beaker. Ultrasound is not usually necessary and
will almost certainly affect the size of liquid particles.
SAMPLE
Analyse
Analyse
Take Representative Sample
(mix well or riffle if dry powder)
Analyse
Ultrasound if
necessary
Try Surfactant
Ultrasound if
necessary
Ultrasound if
necessary
e.g. Ethanol
Propan-2-ol (IPA)
Methanol
Acetone
Butanone (Methyl Ethyl
Ketone)
Hexane
Tolu e ne
Dimethyl Digol
Try Solvent
Does it
disperse in
water?
Does it
float?
Does
Ultrasound
work?
Does this
disperse it?
Analyse
Yes No
No
No
No
Yes
Yes
Yes

Table of Contents

Other manuals for Malvern Mastersizer 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Malvern Mastersizer 3000 and is the answer not in the manual?

Malvern Mastersizer 3000 Specifications

General IconGeneral
Measurement principleLaser diffraction
Blue laser wavelength470 nm
Sample dispersion optionsWet and dry dispersion
Data acquisition rate10 kHz
Particle size range0.01 to 3500 microns
Laser typeHe-Ne and solid state
AccuracyBetter than 1%
RepeatabilityBetter than 0.5%
TechnologyMie and Fraunhofer theories
Sample dispersion unitsHydro MV, Hydro LV, Aero S
Optical modelMie theory
Power requirements100-240 V, 50/60 Hz
Detector arrangementLog-spaced array