EasyManuals Logo

Malvern Mastersizer 3000 User Manual

Malvern Mastersizer 3000
182 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #97 background imageLoading...
Page #97 background image
Making measurements Chapter 4
Mastersizer 3000 Page 4-43
Advanced button
Analysis Model -
Mastersizer
2000/2000E
emulator mode
(continued)
Single mode (Emulated MS2000 / MS2000E) – this
model is appropriate for sample consisting of one very nar-
row mode, such as polystyrene lattices.
Fine Powder
Mode
(Dry dispersion
units only)
This option should be enabled for samples containing a sig-
nificant proportion of material below 10 microns in size. It
can help improve the result reproducibility, and can be used
with any of the Mastersizer 3000 analysis modes.
Section/option Description

Table of Contents

Other manuals for Malvern Mastersizer 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Malvern Mastersizer 3000 and is the answer not in the manual?

Malvern Mastersizer 3000 Specifications

General IconGeneral
Measurement principleLaser diffraction
Blue laser wavelength470 nm
Sample dispersion optionsWet and dry dispersion
Data acquisition rate10 kHz
Particle size range0.01 to 3500 microns
Laser typeHe-Ne and solid state
AccuracyBetter than 1%
RepeatabilityBetter than 0.5%
TechnologyMie and Fraunhofer theories
Sample dispersion unitsHydro MV, Hydro LV, Aero S
Optical modelMie theory
Power requirements100-240 V, 50/60 Hz
Detector arrangementLog-spaced array