Chapter 4 Making measurements
Page 4-40 MAN 0474
Ultrasonication
(Wet dispersion
units only)
Ultrasound can reduce agglomeration, which may help fur-
ther with cleaning. Select whether to Use ultrasound during
the clean sequence or No ultrasound during clean.
Air pressure
(Dry dispersion
units only)
Use the slider bar to set the air pressure at which any
remaining sample is removed (from 0-4 bar, in 0.1 bar incre-
ments).
Feed rate
(Dry dispersion
units only)
Use the slider bar to set the rate at which the remaining sam-
ple is removed - effectively this controls the vibration speed
of the feed tray.
Section/option Description

Table of Contents

Other manuals for Malvern Mastersizer 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Malvern Mastersizer 3000 and is the answer not in the manual?

Malvern Mastersizer 3000 Specifications

General IconGeneral
Measurement principleLaser diffraction
Blue laser wavelength470 nm
Sample dispersion optionsWet and dry dispersion
Data acquisition rate10 kHz
Particle size range0.01 to 3500 microns
Laser typeHe-Ne and solid state
AccuracyBetter than 1%
RepeatabilityBetter than 0.5%
TechnologyMie and Fraunhofer theories
Sample dispersion unitsHydro MV, Hydro LV, Aero S
Optical modelMie theory
Power requirements100-240 V, 50/60 Hz
Detector arrangementLog-spaced array

Related product manuals