79
Measurement
5
5.4. Measurement in sequential mode
2) Set [WAITING TIME FOR JUDGE. ]
1. Enter [WAITING TIME FOR JUDGE. ].
$OORZDEOHUDQJHWR
Default: 0.00
POINT
- Refer to "3.5. Input of numerical values"
IRUPRUHLQIRUPDWLRQDERXWKRZWRHQWHU
numerical values.
2. Tap [OK].
[WAITING TIME FOR JUDGE. ] has been set, and
>6(48(1&(&21752/@ZLOOEHGLVSOD\HG
&RQ¿UPWKHVHWWLQJV
&RQ¿UPWKHYDOXHRI>:$,7,1*7,0()25
JUDGE. ].
The setting is completed if the value set by the step
2) appears under [WAITING TIME FOR JUDGE. ].
(8) Settings of [JUDGE CONDITION] on completion of measurement
Set the condition to judge the measurement as [OK], [OVER],[UNDER], and etc. after
completion of the measurement
1) Display the screen to set [JUDGE CONDITION]
1. Tap [JUDGE CONDITION] on [SEQUENCE
CONTROL 1/3].
7KHVFUHHQWRVHW>-8'*(&21',7,21@ZLOOEH
displayed.
2) Set [JUDGE CONDITION]
1. Select [JUDGE CONDITION].
[TIMER IS UP]0DNHMXGJPHQWZKHQWKHWLPHVHW
by [TIMER IS UP] elapses.
[STAB. AND TIMER IS UP]0DNHMXGJPHQWZKHQ
the time set by [TIMER IS UP] elapses and the
ZHLJKLQJYDOXHEHFRPHVVWDEOH
[STAB. OR TIMER IS UP]0DNHMXGJPHQWZKHQ
the time set by [TIMER IS UP] elapses or the
ZHLJKLQJYDOXHEHFRPHVVWDEOH
Default: [STAB. AND TIMER IS UP]