101 / 170
Possible reasons are:
The integration time is too low;
The selected activity is higher than the actual detected activity.
If statistics are missing, refer to the following instructions:
Do one of the following:
- Lower the activity.
- Set a longer integration time.
- Let the “Auto” function estimate the necessary time
Start a new measurement.
When the measurement is complete, the conversion coefficient “Calculate button of conversion
coefficient calculation is available
8. Click on “Calculate” button.
After calculation,
The conversion coefficient is displayed,
The “Calculation” button is disabled,
The “Write” button is enabled.
9. Click on the button "Write" to enter in red the coefficient in the window of "Calibration" gamma